Apogee Flow Systems Ltd auto40 flow cytometer
Gating strategy for volumetric, CD45-assisted panleucogating <t>Auto40</t> flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK), as depicted from typical figure provided by “Auto-Lymphocyte” software (Apogee Flow Systems Ltd) for the counting of CD4 T lymphocytes: A . CD45-PE-Dyomics649 x axis-light scatter dot plot scattergram; the primary CD45 gating differentiates clearly the CD45-positive lymphocytes and monocyte populations, and the population of polymorphonuclear cells; B . CD4-PE x axis-light scatter dot plot scattergram; double secondary CD4 gatings allow counting independently the population of CD4-negative lymphocyte and CD4-positive lymphocytes; C . Representative figure depicted by the software “Auto-Lymphocyte” in one blood sample taken as example: ( a ) CD45 (FL2)xside scatter (LS2) dot plot scattergram; ( b ) CD4 (FL1)xside scatter (LS2) dot plot scattergram; ( c ) Distribution of bright CD45 events during blood sample counting; ( d ) Distribution of bright CD4 events within the primary CD45 gating scattergram, during blood sample counting.
Auto40 Flow Cytometer, supplied by Apogee Flow Systems Ltd, used in various techniques. Bioz Stars score: 90/100, based on 1 PubMed citations. ZERO BIAS - scores, article reviews, protocol conditions and more
https://www.bioz.com/result/auto40 flow cytometer/product/Apogee Flow Systems Ltd
Average 90 stars, based on 1 article reviews
auto40 flow cytometer - by Bioz Stars, 2026-06
90/100 stars

Images

1) Product Images from "Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resource-constrained settings"

Article Title: Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resource-constrained settings

Journal: BMC Research Notes

doi: 10.1186/1756-0500-6-169

Gating strategy for volumetric, CD45-assisted panleucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK), as depicted from typical figure provided by “Auto-Lymphocyte” software (Apogee Flow Systems Ltd) for the counting of CD4 T lymphocytes: A . CD45-PE-Dyomics649 x axis-light scatter dot plot scattergram; the primary CD45 gating differentiates clearly the CD45-positive lymphocytes and monocyte populations, and the population of polymorphonuclear cells; B . CD4-PE x axis-light scatter dot plot scattergram; double secondary CD4 gatings allow counting independently the population of CD4-negative lymphocyte and CD4-positive lymphocytes; C . Representative figure depicted by the software “Auto-Lymphocyte” in one blood sample taken as example: ( a ) CD45 (FL2)xside scatter (LS2) dot plot scattergram; ( b ) CD4 (FL1)xside scatter (LS2) dot plot scattergram; ( c ) Distribution of bright CD45 events during blood sample counting; ( d ) Distribution of bright CD4 events within the primary CD45 gating scattergram, during blood sample counting.
Figure Legend Snippet: Gating strategy for volumetric, CD45-assisted panleucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK), as depicted from typical figure provided by “Auto-Lymphocyte” software (Apogee Flow Systems Ltd) for the counting of CD4 T lymphocytes: A . CD45-PE-Dyomics649 x axis-light scatter dot plot scattergram; the primary CD45 gating differentiates clearly the CD45-positive lymphocytes and monocyte populations, and the population of polymorphonuclear cells; B . CD4-PE x axis-light scatter dot plot scattergram; double secondary CD4 gatings allow counting independently the population of CD4-negative lymphocyte and CD4-positive lymphocytes; C . Representative figure depicted by the software “Auto-Lymphocyte” in one blood sample taken as example: ( a ) CD45 (FL2)xside scatter (LS2) dot plot scattergram; ( b ) CD4 (FL1)xside scatter (LS2) dot plot scattergram; ( c ) Distribution of bright CD45 events during blood sample counting; ( d ) Distribution of bright CD4 events within the primary CD45 gating scattergram, during blood sample counting.

Techniques Used: Flow Cytometry, Software

CD4 T cell counting in absolute cells/μl and percentage in 182 HIV-1-infected adults and 76 HIV-1-infected children older than 5 years, by the Apogee  Auto40 flow cytometer  at the hôpital Militaire D’Instruction, N’Djamena, Chad, and by the FACSCalibur at the hôpital de la Liberté, N’Djamena
Figure Legend Snippet: CD4 T cell counting in absolute cells/μl and percentage in 182 HIV-1-infected adults and 76 HIV-1-infected children older than 5 years, by the Apogee Auto40 flow cytometer at the hôpital Militaire D’Instruction, N’Djamena, Chad, and by the FACSCalibur at the hôpital de la Liberté, N’Djamena

Techniques Used: Cell Counting, Flow Cytometry

Results from CD4 T cell count measurements in 182 HIV-1-infected adults and 76 HIV-1-infected children older than 5 years, expressed in absolute number (cells/μl) and in percentage carried out in parallel by the Auto40 flow cytometer and by the FACSCalibur. A , A ’, C and C ’. Passing-Bablok agreement tests between the CD4 T cell count results obtained by Auto40 and FACSCalibur flow cytometers, in absolute number ( A and A’ ) and in percentage ( C and C’ ). The diagonal dotted line represents the ideal line (no bias). The full line represents the regression line of the distribution; B , B ’, D and D ’. Bland-Altman analyses on the relative differences between the CD4 T cell counts obtained by Auto40 and FACSCalibur flow cytometers compared with the average CD4 T cell count, in absolute number ( B and B’ ) and in percentage ( D and D’ ). The full line represents the mean relative difference, and the dotted lines represent the superior and inferior limits of agreement. The arrow corresponds to the x abscise axis.
Figure Legend Snippet: Results from CD4 T cell count measurements in 182 HIV-1-infected adults and 76 HIV-1-infected children older than 5 years, expressed in absolute number (cells/μl) and in percentage carried out in parallel by the Auto40 flow cytometer and by the FACSCalibur. A , A ’, C and C ’. Passing-Bablok agreement tests between the CD4 T cell count results obtained by Auto40 and FACSCalibur flow cytometers, in absolute number ( A and A’ ) and in percentage ( C and C’ ). The diagonal dotted line represents the ideal line (no bias). The full line represents the regression line of the distribution; B , B ’, D and D ’. Bland-Altman analyses on the relative differences between the CD4 T cell counts obtained by Auto40 and FACSCalibur flow cytometers compared with the average CD4 T cell count, in absolute number ( B and B’ ) and in percentage ( D and D’ ). The full line represents the mean relative difference, and the dotted lines represent the superior and inferior limits of agreement. The arrow corresponds to the x abscise axis.

Techniques Used: Cell Counting, Infection, Flow Cytometry

The sensitivity and specificity of CD4 T cell counting by the  Auto40  to identify patients having less than (or more than) 200 CD4 T cells/μl, 350 CD4 T cells/μl, 750 CD4 T cells/μl and 25%CD4+, calculated on the 258 available CD4 T cell count measurements
Figure Legend Snippet: The sensitivity and specificity of CD4 T cell counting by the Auto40 to identify patients having less than (or more than) 200 CD4 T cells/μl, 350 CD4 T cells/μl, 750 CD4 T cells/μl and 25%CD4+, calculated on the 258 available CD4 T cell count measurements

Techniques Used: Cell Counting

CD4 T cell counting in absolute count and percentage in 258 HIV-1-infected individuals, by the Apogee  Auto40 flow cytometer  at the hôpital Militaire D’Instruction, N’Djamena, Chad, and by the FACSCalibur at the hôpital de la Liberté, N’Djamena, at various CD4 T cell count ranges according to the FACSCalibur results
Figure Legend Snippet: CD4 T cell counting in absolute count and percentage in 258 HIV-1-infected individuals, by the Apogee Auto40 flow cytometer at the hôpital Militaire D’Instruction, N’Djamena, Chad, and by the FACSCalibur at the hôpital de la Liberté, N’Djamena, at various CD4 T cell count ranges according to the FACSCalibur results

Techniques Used: Cell Counting, Flow Cytometry



Similar Products

90
Apogee Flow Systems Ltd auto40 flow cytometer
Gating strategy for volumetric, CD45-assisted panleucogating <t>Auto40</t> flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK), as depicted from typical figure provided by “Auto-Lymphocyte” software (Apogee Flow Systems Ltd) for the counting of CD4 T lymphocytes: A . CD45-PE-Dyomics649 x axis-light scatter dot plot scattergram; the primary CD45 gating differentiates clearly the CD45-positive lymphocytes and monocyte populations, and the population of polymorphonuclear cells; B . CD4-PE x axis-light scatter dot plot scattergram; double secondary CD4 gatings allow counting independently the population of CD4-negative lymphocyte and CD4-positive lymphocytes; C . Representative figure depicted by the software “Auto-Lymphocyte” in one blood sample taken as example: ( a ) CD45 (FL2)xside scatter (LS2) dot plot scattergram; ( b ) CD4 (FL1)xside scatter (LS2) dot plot scattergram; ( c ) Distribution of bright CD45 events during blood sample counting; ( d ) Distribution of bright CD4 events within the primary CD45 gating scattergram, during blood sample counting.
Auto40 Flow Cytometer, supplied by Apogee Flow Systems Ltd, used in various techniques. Bioz Stars score: 90/100, based on 1 PubMed citations. ZERO BIAS - scores, article reviews, protocol conditions and more
https://www.bioz.com/result/auto40 flow cytometer/product/Apogee Flow Systems Ltd
Average 90 stars, based on 1 article reviews
auto40 flow cytometer - by Bioz Stars, 2026-06
90/100 stars
  Buy from Supplier

90
Apogee Flow Systems Ltd auto40
Gating strategy for volumetric, CD45-assisted panleucogating <t>Auto40</t> flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK), as depicted from typical figure provided by “Auto-Lymphocyte” software (Apogee Flow Systems Ltd) for the counting of CD4 T lymphocytes: A . CD45-PE-Dyomics649 x axis-light scatter dot plot scattergram; the primary CD45 gating differentiates clearly the CD45-positive lymphocytes and monocyte populations, and the population of polymorphonuclear cells; B . CD4-PE x axis-light scatter dot plot scattergram; double secondary CD4 gatings allow counting independently the population of CD4-negative lymphocyte and CD4-positive lymphocytes; C . Representative figure depicted by the software “Auto-Lymphocyte” in one blood sample taken as example: ( a ) CD45 (FL2)xside scatter (LS2) dot plot scattergram; ( b ) CD4 (FL1)xside scatter (LS2) dot plot scattergram; ( c ) Distribution of bright CD45 events during blood sample counting; ( d ) Distribution of bright CD4 events within the primary CD45 gating scattergram, during blood sample counting.
Auto40, supplied by Apogee Flow Systems Ltd, used in various techniques. Bioz Stars score: 90/100, based on 1 PubMed citations. ZERO BIAS - scores, article reviews, protocol conditions and more
https://www.bioz.com/result/auto40/product/Apogee Flow Systems Ltd
Average 90 stars, based on 1 article reviews
auto40 - by Bioz Stars, 2026-06
90/100 stars
  Buy from Supplier

90
Apogee Flow Systems Ltd flow cytometer auto40
Gating strategy for volumetric, CD45-assisted panleucogating <t>Auto40</t> flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK), as depicted from typical figure provided by “Auto-Lymphocyte” software (Apogee Flow Systems Ltd) for the counting of CD4 T lymphocytes: A . CD45-PE-Dyomics649 x axis-light scatter dot plot scattergram; the primary CD45 gating differentiates clearly the CD45-positive lymphocytes and monocyte populations, and the population of polymorphonuclear cells; B . CD4-PE x axis-light scatter dot plot scattergram; double secondary CD4 gatings allow counting independently the population of CD4-negative lymphocyte and CD4-positive lymphocytes; C . Representative figure depicted by the software “Auto-Lymphocyte” in one blood sample taken as example: ( a ) CD45 (FL2)xside scatter (LS2) dot plot scattergram; ( b ) CD4 (FL1)xside scatter (LS2) dot plot scattergram; ( c ) Distribution of bright CD45 events during blood sample counting; ( d ) Distribution of bright CD4 events within the primary CD45 gating scattergram, during blood sample counting.
Flow Cytometer Auto40, supplied by Apogee Flow Systems Ltd, used in various techniques. Bioz Stars score: 90/100, based on 1 PubMed citations. ZERO BIAS - scores, article reviews, protocol conditions and more
https://www.bioz.com/result/flow cytometer auto40/product/Apogee Flow Systems Ltd
Average 90 stars, based on 1 article reviews
flow cytometer auto40 - by Bioz Stars, 2026-06
90/100 stars
  Buy from Supplier

90
Apogee Flow Systems Ltd calibrated bead sample apogee cat # 1444 for auto40-green
Gating strategy for volumetric, CD45-assisted panleucogating <t>Auto40</t> flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK), as depicted from typical figure provided by “Auto-Lymphocyte” software (Apogee Flow Systems Ltd) for the counting of CD4 T lymphocytes: A . CD45-PE-Dyomics649 x axis-light scatter dot plot scattergram; the primary CD45 gating differentiates clearly the CD45-positive lymphocytes and monocyte populations, and the population of polymorphonuclear cells; B . CD4-PE x axis-light scatter dot plot scattergram; double secondary CD4 gatings allow counting independently the population of CD4-negative lymphocyte and CD4-positive lymphocytes; C . Representative figure depicted by the software “Auto-Lymphocyte” in one blood sample taken as example: ( a ) CD45 (FL2)xside scatter (LS2) dot plot scattergram; ( b ) CD4 (FL1)xside scatter (LS2) dot plot scattergram; ( c ) Distribution of bright CD45 events during blood sample counting; ( d ) Distribution of bright CD4 events within the primary CD45 gating scattergram, during blood sample counting.
Calibrated Bead Sample Apogee Cat # 1444 For Auto40 Green, supplied by Apogee Flow Systems Ltd, used in various techniques. Bioz Stars score: 90/100, based on 1 PubMed citations. ZERO BIAS - scores, article reviews, protocol conditions and more
https://www.bioz.com/result/calibrated bead sample apogee cat # 1444 for auto40-green/product/Apogee Flow Systems Ltd
Average 90 stars, based on 1 article reviews
calibrated bead sample apogee cat # 1444 for auto40-green - by Bioz Stars, 2026-06
90/100 stars
  Buy from Supplier

Image Search Results


Gating strategy for volumetric, CD45-assisted panleucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK), as depicted from typical figure provided by “Auto-Lymphocyte” software (Apogee Flow Systems Ltd) for the counting of CD4 T lymphocytes: A . CD45-PE-Dyomics649 x axis-light scatter dot plot scattergram; the primary CD45 gating differentiates clearly the CD45-positive lymphocytes and monocyte populations, and the population of polymorphonuclear cells; B . CD4-PE x axis-light scatter dot plot scattergram; double secondary CD4 gatings allow counting independently the population of CD4-negative lymphocyte and CD4-positive lymphocytes; C . Representative figure depicted by the software “Auto-Lymphocyte” in one blood sample taken as example: ( a ) CD45 (FL2)xside scatter (LS2) dot plot scattergram; ( b ) CD4 (FL1)xside scatter (LS2) dot plot scattergram; ( c ) Distribution of bright CD45 events during blood sample counting; ( d ) Distribution of bright CD4 events within the primary CD45 gating scattergram, during blood sample counting.

Journal: BMC Research Notes

Article Title: Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resource-constrained settings

doi: 10.1186/1756-0500-6-169

Figure Lengend Snippet: Gating strategy for volumetric, CD45-assisted panleucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK), as depicted from typical figure provided by “Auto-Lymphocyte” software (Apogee Flow Systems Ltd) for the counting of CD4 T lymphocytes: A . CD45-PE-Dyomics649 x axis-light scatter dot plot scattergram; the primary CD45 gating differentiates clearly the CD45-positive lymphocytes and monocyte populations, and the population of polymorphonuclear cells; B . CD4-PE x axis-light scatter dot plot scattergram; double secondary CD4 gatings allow counting independently the population of CD4-negative lymphocyte and CD4-positive lymphocytes; C . Representative figure depicted by the software “Auto-Lymphocyte” in one blood sample taken as example: ( a ) CD45 (FL2)xside scatter (LS2) dot plot scattergram; ( b ) CD4 (FL1)xside scatter (LS2) dot plot scattergram; ( c ) Distribution of bright CD45 events during blood sample counting; ( d ) Distribution of bright CD4 events within the primary CD45 gating scattergram, during blood sample counting.

Article Snippet: The absolute and percentage CD4 T cell counts of 258 HIV-1-infected blood samples (182 adults and 76 children), living in N’Djamena, Chad, were performed by single-platform, volumetric, CD45-assisted pan-leucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK) comparing to the FACSCalibur flow cytometer as a reference method.

Techniques: Flow Cytometry, Software

CD4 T cell counting in absolute cells/μl and percentage in 182 HIV-1-infected adults and 76 HIV-1-infected children older than 5 years, by the Apogee  Auto40 flow cytometer  at the hôpital Militaire D’Instruction, N’Djamena, Chad, and by the FACSCalibur at the hôpital de la Liberté, N’Djamena

Journal: BMC Research Notes

Article Title: Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resource-constrained settings

doi: 10.1186/1756-0500-6-169

Figure Lengend Snippet: CD4 T cell counting in absolute cells/μl and percentage in 182 HIV-1-infected adults and 76 HIV-1-infected children older than 5 years, by the Apogee Auto40 flow cytometer at the hôpital Militaire D’Instruction, N’Djamena, Chad, and by the FACSCalibur at the hôpital de la Liberté, N’Djamena

Article Snippet: The absolute and percentage CD4 T cell counts of 258 HIV-1-infected blood samples (182 adults and 76 children), living in N’Djamena, Chad, were performed by single-platform, volumetric, CD45-assisted pan-leucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK) comparing to the FACSCalibur flow cytometer as a reference method.

Techniques: Cell Counting, Flow Cytometry

Results from CD4 T cell count measurements in 182 HIV-1-infected adults and 76 HIV-1-infected children older than 5 years, expressed in absolute number (cells/μl) and in percentage carried out in parallel by the Auto40 flow cytometer and by the FACSCalibur. A , A ’, C and C ’. Passing-Bablok agreement tests between the CD4 T cell count results obtained by Auto40 and FACSCalibur flow cytometers, in absolute number ( A and A’ ) and in percentage ( C and C’ ). The diagonal dotted line represents the ideal line (no bias). The full line represents the regression line of the distribution; B , B ’, D and D ’. Bland-Altman analyses on the relative differences between the CD4 T cell counts obtained by Auto40 and FACSCalibur flow cytometers compared with the average CD4 T cell count, in absolute number ( B and B’ ) and in percentage ( D and D’ ). The full line represents the mean relative difference, and the dotted lines represent the superior and inferior limits of agreement. The arrow corresponds to the x abscise axis.

Journal: BMC Research Notes

Article Title: Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resource-constrained settings

doi: 10.1186/1756-0500-6-169

Figure Lengend Snippet: Results from CD4 T cell count measurements in 182 HIV-1-infected adults and 76 HIV-1-infected children older than 5 years, expressed in absolute number (cells/μl) and in percentage carried out in parallel by the Auto40 flow cytometer and by the FACSCalibur. A , A ’, C and C ’. Passing-Bablok agreement tests between the CD4 T cell count results obtained by Auto40 and FACSCalibur flow cytometers, in absolute number ( A and A’ ) and in percentage ( C and C’ ). The diagonal dotted line represents the ideal line (no bias). The full line represents the regression line of the distribution; B , B ’, D and D ’. Bland-Altman analyses on the relative differences between the CD4 T cell counts obtained by Auto40 and FACSCalibur flow cytometers compared with the average CD4 T cell count, in absolute number ( B and B’ ) and in percentage ( D and D’ ). The full line represents the mean relative difference, and the dotted lines represent the superior and inferior limits of agreement. The arrow corresponds to the x abscise axis.

Article Snippet: The absolute and percentage CD4 T cell counts of 258 HIV-1-infected blood samples (182 adults and 76 children), living in N’Djamena, Chad, were performed by single-platform, volumetric, CD45-assisted pan-leucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK) comparing to the FACSCalibur flow cytometer as a reference method.

Techniques: Cell Counting, Infection, Flow Cytometry

The sensitivity and specificity of CD4 T cell counting by the  Auto40  to identify patients having less than (or more than) 200 CD4 T cells/μl, 350 CD4 T cells/μl, 750 CD4 T cells/μl and 25%CD4+, calculated on the 258 available CD4 T cell count measurements

Journal: BMC Research Notes

Article Title: Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resource-constrained settings

doi: 10.1186/1756-0500-6-169

Figure Lengend Snippet: The sensitivity and specificity of CD4 T cell counting by the Auto40 to identify patients having less than (or more than) 200 CD4 T cells/μl, 350 CD4 T cells/μl, 750 CD4 T cells/μl and 25%CD4+, calculated on the 258 available CD4 T cell count measurements

Article Snippet: The absolute and percentage CD4 T cell counts of 258 HIV-1-infected blood samples (182 adults and 76 children), living in N’Djamena, Chad, were performed by single-platform, volumetric, CD45-assisted pan-leucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK) comparing to the FACSCalibur flow cytometer as a reference method.

Techniques: Cell Counting

CD4 T cell counting in absolute count and percentage in 258 HIV-1-infected individuals, by the Apogee  Auto40 flow cytometer  at the hôpital Militaire D’Instruction, N’Djamena, Chad, and by the FACSCalibur at the hôpital de la Liberté, N’Djamena, at various CD4 T cell count ranges according to the FACSCalibur results

Journal: BMC Research Notes

Article Title: Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resource-constrained settings

doi: 10.1186/1756-0500-6-169

Figure Lengend Snippet: CD4 T cell counting in absolute count and percentage in 258 HIV-1-infected individuals, by the Apogee Auto40 flow cytometer at the hôpital Militaire D’Instruction, N’Djamena, Chad, and by the FACSCalibur at the hôpital de la Liberté, N’Djamena, at various CD4 T cell count ranges according to the FACSCalibur results

Article Snippet: The absolute and percentage CD4 T cell counts of 258 HIV-1-infected blood samples (182 adults and 76 children), living in N’Djamena, Chad, were performed by single-platform, volumetric, CD45-assisted pan-leucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK) comparing to the FACSCalibur flow cytometer as a reference method.

Techniques: Cell Counting, Flow Cytometry

Gating strategy for volumetric, CD45-assisted panleucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK), as depicted from typical figure provided by “Auto-Lymphocyte” software (Apogee Flow Systems Ltd) for the counting of CD4 T lymphocytes: A . CD45-PE-Dyomics649 x axis-light scatter dot plot scattergram; the primary CD45 gating differentiates clearly the CD45-positive lymphocytes and monocyte populations, and the population of polymorphonuclear cells; B . CD4-PE x axis-light scatter dot plot scattergram; double secondary CD4 gatings allow counting independently the population of CD4-negative lymphocyte and CD4-positive lymphocytes; C . Representative figure depicted by the software “Auto-Lymphocyte” in one blood sample taken as example: ( a ) CD45 (FL2)xside scatter (LS2) dot plot scattergram; ( b ) CD4 (FL1)xside scatter (LS2) dot plot scattergram; ( c ) Distribution of bright CD45 events during blood sample counting; ( d ) Distribution of bright CD4 events within the primary CD45 gating scattergram, during blood sample counting.

Journal: BMC Research Notes

Article Title: Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resource-constrained settings

doi: 10.1186/1756-0500-6-169

Figure Lengend Snippet: Gating strategy for volumetric, CD45-assisted panleucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK), as depicted from typical figure provided by “Auto-Lymphocyte” software (Apogee Flow Systems Ltd) for the counting of CD4 T lymphocytes: A . CD45-PE-Dyomics649 x axis-light scatter dot plot scattergram; the primary CD45 gating differentiates clearly the CD45-positive lymphocytes and monocyte populations, and the population of polymorphonuclear cells; B . CD4-PE x axis-light scatter dot plot scattergram; double secondary CD4 gatings allow counting independently the population of CD4-negative lymphocyte and CD4-positive lymphocytes; C . Representative figure depicted by the software “Auto-Lymphocyte” in one blood sample taken as example: ( a ) CD45 (FL2)xside scatter (LS2) dot plot scattergram; ( b ) CD4 (FL1)xside scatter (LS2) dot plot scattergram; ( c ) Distribution of bright CD45 events during blood sample counting; ( d ) Distribution of bright CD4 events within the primary CD45 gating scattergram, during blood sample counting.

Article Snippet: Direct volumetric CD4 T cell measurements were performed on the Auto40 using phycoerythrin (PE)-conjugated anti-CD4 and PE-Dyomics649-conjugated anti-CD45 monoclonal antibodies (Apogee Flow Systems Ltd).

Techniques: Flow Cytometry, Software

CD4 T cell counting in absolute cells/μl and percentage in 182 HIV-1-infected adults and 76 HIV-1-infected children older than 5 years, by the Apogee  Auto40  flow cytometer at the hôpital Militaire D’Instruction, N’Djamena, Chad, and by the FACSCalibur at the hôpital de la Liberté, N’Djamena

Journal: BMC Research Notes

Article Title: Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resource-constrained settings

doi: 10.1186/1756-0500-6-169

Figure Lengend Snippet: CD4 T cell counting in absolute cells/μl and percentage in 182 HIV-1-infected adults and 76 HIV-1-infected children older than 5 years, by the Apogee Auto40 flow cytometer at the hôpital Militaire D’Instruction, N’Djamena, Chad, and by the FACSCalibur at the hôpital de la Liberté, N’Djamena

Article Snippet: Direct volumetric CD4 T cell measurements were performed on the Auto40 using phycoerythrin (PE)-conjugated anti-CD4 and PE-Dyomics649-conjugated anti-CD45 monoclonal antibodies (Apogee Flow Systems Ltd).

Techniques: Cell Counting, Flow Cytometry

Results from CD4 T cell count measurements in 182 HIV-1-infected adults and 76 HIV-1-infected children older than 5 years, expressed in absolute number (cells/μl) and in percentage carried out in parallel by the Auto40 flow cytometer and by the FACSCalibur. A , A ’, C and C ’. Passing-Bablok agreement tests between the CD4 T cell count results obtained by Auto40 and FACSCalibur flow cytometers, in absolute number ( A and A’ ) and in percentage ( C and C’ ). The diagonal dotted line represents the ideal line (no bias). The full line represents the regression line of the distribution; B , B ’, D and D ’. Bland-Altman analyses on the relative differences between the CD4 T cell counts obtained by Auto40 and FACSCalibur flow cytometers compared with the average CD4 T cell count, in absolute number ( B and B’ ) and in percentage ( D and D’ ). The full line represents the mean relative difference, and the dotted lines represent the superior and inferior limits of agreement. The arrow corresponds to the x abscise axis.

Journal: BMC Research Notes

Article Title: Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resource-constrained settings

doi: 10.1186/1756-0500-6-169

Figure Lengend Snippet: Results from CD4 T cell count measurements in 182 HIV-1-infected adults and 76 HIV-1-infected children older than 5 years, expressed in absolute number (cells/μl) and in percentage carried out in parallel by the Auto40 flow cytometer and by the FACSCalibur. A , A ’, C and C ’. Passing-Bablok agreement tests between the CD4 T cell count results obtained by Auto40 and FACSCalibur flow cytometers, in absolute number ( A and A’ ) and in percentage ( C and C’ ). The diagonal dotted line represents the ideal line (no bias). The full line represents the regression line of the distribution; B , B ’, D and D ’. Bland-Altman analyses on the relative differences between the CD4 T cell counts obtained by Auto40 and FACSCalibur flow cytometers compared with the average CD4 T cell count, in absolute number ( B and B’ ) and in percentage ( D and D’ ). The full line represents the mean relative difference, and the dotted lines represent the superior and inferior limits of agreement. The arrow corresponds to the x abscise axis.

Article Snippet: Direct volumetric CD4 T cell measurements were performed on the Auto40 using phycoerythrin (PE)-conjugated anti-CD4 and PE-Dyomics649-conjugated anti-CD45 monoclonal antibodies (Apogee Flow Systems Ltd).

Techniques: Cell Counting, Infection, Flow Cytometry

The sensitivity and specificity of CD4 T cell counting by the  Auto40  to identify patients having less than (or more than) 200 CD4 T cells/μl, 350 CD4 T cells/μl, 750 CD4 T cells/μl and 25%CD4+, calculated on the 258 available CD4 T cell count measurements

Journal: BMC Research Notes

Article Title: Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resource-constrained settings

doi: 10.1186/1756-0500-6-169

Figure Lengend Snippet: The sensitivity and specificity of CD4 T cell counting by the Auto40 to identify patients having less than (or more than) 200 CD4 T cells/μl, 350 CD4 T cells/μl, 750 CD4 T cells/μl and 25%CD4+, calculated on the 258 available CD4 T cell count measurements

Article Snippet: Direct volumetric CD4 T cell measurements were performed on the Auto40 using phycoerythrin (PE)-conjugated anti-CD4 and PE-Dyomics649-conjugated anti-CD45 monoclonal antibodies (Apogee Flow Systems Ltd).

Techniques: Cell Counting

CD4 T cell counting in absolute count and percentage in 258 HIV-1-infected individuals, by the Apogee  Auto40  flow cytometer at the hôpital Militaire D’Instruction, N’Djamena, Chad, and by the FACSCalibur at the hôpital de la Liberté, N’Djamena, at various CD4 T cell count ranges according to the FACSCalibur results

Journal: BMC Research Notes

Article Title: Single-platform, volumetric, CD45-assisted pan-leucogating flow cytometry for CD4 T lymphocytes monitoring of HIV infection according to the WHO recommendations for resource-constrained settings

doi: 10.1186/1756-0500-6-169

Figure Lengend Snippet: CD4 T cell counting in absolute count and percentage in 258 HIV-1-infected individuals, by the Apogee Auto40 flow cytometer at the hôpital Militaire D’Instruction, N’Djamena, Chad, and by the FACSCalibur at the hôpital de la Liberté, N’Djamena, at various CD4 T cell count ranges according to the FACSCalibur results

Article Snippet: Direct volumetric CD4 T cell measurements were performed on the Auto40 using phycoerythrin (PE)-conjugated anti-CD4 and PE-Dyomics649-conjugated anti-CD45 monoclonal antibodies (Apogee Flow Systems Ltd).

Techniques: Cell Counting, Flow Cytometry